Temperature Monitoring & Alarm Unit

Thermographic Printer Channel
Record and print simultaneously three different temperatures of -20 to 85 ° C and with time and date printing.
Informs and warns of a possible increase or decrease in temperature beyond the limits you set.
Protect your products and your customers from amfivolos preserved foods with 24-hour surveillance 24 hours a day, 365 days a year.
Printing cost only 1 to 2 per month.

Overview

General features of the device:
The device is a metal box 20x20x8 cm suitable for fixing on the wall.
Testator display 2 lines of 16 characters, only has 4 softkeys planning and control, and integrated printer, 57mm width and 10m in length to 20 characters printed.

During the rendering process risk (Alarm)
the light at the top will blink as the backlight.
The pazer (Buzzer) will sound intermittently and the screen will display a more sensor is the problem though is High or Low Alarm.

Specification

How to schedule and hours:
How easy replacement printing papers
With a flick of a pen you can in 10/2 to place
new roll of paper that costs less than one euro.

Setting and changing hours & days. :
1 - Press the OK button and holding switch
device. When the screen appears (24Hr)
release OK.
2-Following the arrows ↑ ↓ or select the time in 24 hours
basis. Then press OK to change the indication to
minutes and then continue until the setup option of
chronology, after pressing the NO for 3 seconds until
be stored in memory and restart the
appliance.

Set the time between each printing & Device Name:
Turn on the device while holding down the NO until the display shows only the cursor (_).
NO release and you will see on your screen indicating time adjustment in minutes (*) next. Adjust the minutes you want between each print and then 'select by pressing the OK button to adjust the number of attribute name (ID) of your device which will appear in print.
Again the OK option (*) returns to the time setting.
When you wish to exit, press NO for 3 seconds until you see the message - All Saved - on your screen.

Factory data reset (Defaults):
Turn the unit down simultaneously having two arrows ↑ + ↓ keys until you see the message Defaults loaded.

Setting limits Alarms:
(High alarm limits):
While it is fully operational keep ↑ + ↓ keys simultaneously for 3 seconds.
The display will change to TH1. With the arrows ↑ or ↓ to select the sensor (channel) you want to set limit for high temperature (High).
Holding down the OK set the arrows ↑ ↓ or temperature for High alarm.
Press NO for 2 seconds to memorize and exit.
(Low alarm limits):
While it is fully operational keep ↑ + ↓ keys simultaneously for 3 seconds.
The display will change to TH1.
Then press the NO and 1 second into the ↓ key will display the setting TL1 (Low Alarm). With the arrows ↑ or ↓ to select the sensor (channel) you want to configure.
Holding down the OK set the arrows ↑ ↓ or temperature for High alarm.
Press NO for 2 seconds to memorize and exit.
Note:
Whenever we have High Alarm (HV) ie the temperature reading is higher than the default then we audible, bright and relay (relay) 1 (H) output.
Whenever we Low Alarm (LV) i.e. the temperature reading is lower than the predetermined then we audible, luminous and the relay (relay) 2 (L) output.
Stored in memory only the data sheet you set last. Ie only the High alarm or just the Low Alarm limits.